TB 9-6625-2332-50
NOTE
When indications specified in this procedure are not within
tolerance, perform the power supply check prior to making
adjustments.
b. Connect TI to a 115 V ac power source.
c. Press POWER key to on and allow at least 15 minutes for TI to stabilize.
a. Performance Check
(1) On TI press RECALL SETUP pushbutton and select STANDARD key on the
crt.
(2) Press TI keys as indicated in (a) thru (c) below.
(a) MORE.
(b) CAL.
(c) EXECUTE.
(4) Press TI keys as indicated in (a) thru (c) below.
(a) MORE.
(b) SELF TEST.
(c) EXECUTE.
(5) Wait for TI to finish self test.
(6) Connect equipment as shown in figure 2 below.