TB 9-6625-2332-50
indication
Termination
Test instrument
Vhigh
X termination
correction
setting
Max
correction factor
Min
factor
indication
15.0
14.845
15.155
10.0
9.895
10.105
5.0
4.945
5.055
(34) Press TI pushbuttons as listed in (a) through (c) below:
(a) Vhigh and enter 0 from data keyboard and press ENTER/HZ.
(b) Vlow and enter 5 from data keyboard and press ENTER/HZ.
(c) TRIGGER SLOPE POSITIVE.
(35) Record multimeter indication in table 7 under actual multimeter reading for
Vlow setting.
(36) Multiply the actual multimeter reading times the termination correction factor
(37) Verify that the value recorded in (36) above is within limits specified for the
Vlow setting.
(38) Set TI to the next Vlow setting listed in table 7 and repeat (35) through (37)
above.
(39) Repeat (38) above for the remaining Vlow settings listed in table 7.
indication
Test instrument
Termination
X termination
correction
Vlow
correction factor
indication
factor
Min
Max
setting
-5.055
-4.945
-5.0
-10.105
-9.895
-10.0
-15.155
-14.845
-15.0
(40) Press DISABLE pushbutton (9215 output module) to off ( red light lit).
(41) Disconnect TI 9215 module OUTPUT from BNC tee adapter.
b. Adjustments. No adjustments can be made.
9