TB 9-6625-2332-50
Test instrument
Digitizing oscilloscope indications
settings
(sec)
Per
Lead
Trail
Risetime
Falltime
setting
setting
setting
Min
Max
Min
Max
1 sec
6.5e-9
6.5e-9
4.9e-9
8.1e-9
4.9e-9
8.1e-9
10e-9
10e-9
7.7e-9
12.3e-9
7.7e-9
12.3e-9
20e-9
20e-9
15.7e-9
24.3e-9
15.7e-9
24.e3-9
100e-9
100e-9
79.7e-9
120.3e-9
79.7e-9
120.3e-9
l-6
le-6
0.799e-6
1.2e-6
0.799e-6
1.2e-6
100 sec
10e-6
10e-6
8e-6
12e-6
8e-6
12e-6
10 msec
100e-6
100e-6
80e-6
120e-6
80e-6
120e-6
1e-3
1e-3
0.8e-3
1.2e-3
0.8e-3
1.2e-3
200 msec
10e-3
10e-3
8e-3
12e-3
8e-3
12e-3
500 msec
100e-3
100e-3
76e-3
114e-3
76e-3
114e-3
(24) Press DISABLE pushbutton (model 9215 output module) to off ( red light lit).
(25) Disconnect cable from TI 9215 module OUTPUT.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Reset digitizing oscilloscope and cal the gain.
(3) Press TI pushbuttons as listed in (a) through (h) below:
(a)
CHANNEL A.
(b)
Vhigh and
enter 2.5 from data keyboard and
press ENTER/HZ.
(c)
Vlow and
enter 0 from the
data keyboard and
press ENTER/HZ.
(d)
Lead and
enter .9 from data keyboard and
press n/GHz.
(e)
Trail and
enter .9 from data keyboard and
press n/GHz.
(f)
Per and
enter 10 from data keyboard and
press n/GHz.
(g)
Wid and
enter 1.6 from data keyboard and
press n/GHz.
(h)
Press DISABLE pushbutton (model 9211 output module) to on (red light
extinguished).
(6) Verify that the digitizing oscilloscope indicates within the limits listed in table
12 for the
Per and
Wid setting.
(7) Set TI to the next Per and Wid settings listed in table 12.
(8) Repeat (5) through (7) above.
(9) Repeat (8) above for the remaining Per and Wid settings listed in table 12.