TB 9-6625-2332-50
Digitizing oscilloscope indications
Test instrument settings
(sec)
Per
Wid
setting
setting
Min
Max
10 nsec
1.6 nsec
1.29e-9
1.91e-9
10 nsec
5 nsec
4.675e-9
5.25e-9
20 nsec
10 nsec
9.62e-9
10.38e-9
200 nsec
100 nsec
99.2e-9
100.8e-9
1 sec
500 nsec
497.2e-9
502.8e-9
4 sec
2 sec
1.99e-6
2.01e-6
20 sec
sec
9.95e-6
10.05e-6
10
99.5e-6
100.5e-6
200 sec
100 sec
2 msec
1
msec
0.995e-3
1.005e-3
20 msec
10
msec
9.95e-3
10.05e-3
200 msec
100 msec
99.5e-3
100.5e-3
(10) Press DISABLE pushbutton (model 9211 output module) to off ( red light lit).
(11) Disconnect cable from TI 9211 module OUTPUT.
(12) Connect cable to TI 9215 module OUTPUT.
(13) Press TI pushbuttons as listed in (a) through (h) below:
(a)
CHANNEL B.
(b)
Vhigh and
enter 4 from data keyboard and
press ENTER/HZ.
(c)
Vlow and
enter 0 from the
data keyboard and
press ENTER/HZ.
(d)
Lead and
enter 5 from data keyboard and
press n/GHz.
(e)
Trail and
enter 5 from data keyboard and
press n/GHz.
(f)
Per and
enter 20 from data keyboard and
press n/GHz.
(g)
Wid and
enter 10 from data keyboard and
press n/GHz.
(h)
Press, DISABLE pushbutton (model 9215 output module) to on (red light
extinguished).
(16) Verify that the digitizing oscilloscope indicates within the limits listed in table
13 for the
Per and
Wid setting.
(17) Set TI to the next Per and Wid settings listed in table 13.
(18) Repeat (15) through (17) above.
(19) Repeat (18) above for the remaining Per and Wid settings listed in table 13.
15