TB 9-6625-2332-50
(28) Repeat (23) and (24) above.
(29) Repeat (24) through (28) above for the remaining Per, Wid and Delay settings
Test instrument
∆T indications
settings
settings
(sec)
Per
Wid
Delay
Time range
Time delay
Min
Max
40 n
10
n
10 n
50e-9
16e-9
8.95e-9
11.05e-9
200 n
175
n
100 n
200e-9
190e-9
98.5e-9
101.5e-9
2e-6
1.8e-6
0.994e-6
1.006e-6
2
1.75
1
20e-6
18e-6
9.949e-6
10.051e-6
20
17.5
10
200e-6
180e-6
99.5e-6
100.5e-6
200
100
175
2m
1.75 m
1m
2e-3
1.8e-3
0.995e-3
1.005e-3
20 m
17.5 m
10 m
20e-3
18e-3
9.95e-3
10.05e-3
200 m
175
m
100 m
200e-3
180e-3
99.5e-3
100.5e-3
(30) Press DISABLE pushbutton (model 9215 output module) to off ( red light lit).
(31) Disconnect cable from TI 9215 module OUTPUT.
b. Adjustments. No adjustments can be made.
14. Trigger Level and Polarity
a. Performance Check
(1) Connect equipment as shown in figure 4 below.
20 CHANGE 2
PIN: 080250-002