TB 9-6625-2332-50
Digitizing oscilloscope indications
Test instrument settings
(sec)
Per
Wid
setting
setting
Min
Max
20
nsec
10
nsec
9.62e-9
10.38e-9
200
nsec
100
nsec
99.25e-9
100.8e-9
500
nsec
497.2e-9
502.8e-9
sec
1
1.99e-6
2.01e-6
sec
sec
4
2
9.95e-6
10.05e-6
sec
sec
20
10
sec
sec
99.5e-6
100.5e-6
200
100
2
msec
1
msec
.995e-3
1.005e-3
20
msec
10
msec
9.95e-3
10.05e-3
200
msec
100
msec
99.5e-3
100.5e-3
(20) Press DISABLE pushbutton (model 9215 output module) to off ( red light lit).
(21) Disconnect cable from TI 9215 module OUTPUT.
b. Adjustments. No adjustments can be made.
a. Performance Check
(2) Reset digitizing oscilloscope and cal the gain.
(3) Press TI pushbuttons as listed in (a) through (f) below:
(a)
CHANNEL A.
(b)
Vhigh and
enter 5 from data keyboard and
press ENTER/HZ.
(c)
Vlow and
enter 0 from the
data keyboard and
press ENTER/HZ.
(d)
Lead and
enter .9 from data keyboard and
press n/GHz.
(e)
Trail and
enter .9 from data keyboard and
press n/GHz.
(f)
Per and
enter 20 from data keyboard and
press n/GHz.
(g)
Wid and
enter 10 from data keyboard and
press n/GHz.
(h)
Press DISABLE pushbutton (model 9211 output module) to on (red light
extinguished).
(5) Setup digitizing oscilloscope to display the leading edge of a single pulse at the
center of the display.
(6) Setup digitizing oscilloscope to take a time histogram with 300 samples at the
mid point of the leading edge of the displayed pulse.
(7) Verify that the digitizing oscilloscope sigma indication is less than the limit