TB 9-6625-2333-24
(11) Set synthesizer/level generator frequency to F2 in table 3 for the current TI
RES BW setting.
(12) Press TI MARKER PEAK SEARCH key. Record ' MKR frequency indication
as actual span in table 4 for the current TI RES BW setting.
Table 4. -3 dB Resolution Bandwidth Accuracy
- 3 dB BW
Test instrument
Span
error limits
RES BW settings
(%)
Ideal
Actual
Measured
Corrected
2 MHz
2 MHz
MHz
-25, +50
r25
1 MHz
1 MHz
MHz
r10
300 kHz
300 kHz
kHz
r10
100 kHz
100 kHz
kHz
r10
30 kHz
30 kHz
kHz
r10
10 kHz
10 kHz
kHz
r10
3 kHz
N/A
N/A
r10
1 kHz
N/A
N/A
r10
300 Hz
N/A
N/A
(13) Press TI INSTRUMENT STATE RECALL key then [RECALL STATE] key
and then [STATE 0] key.
(14) Set synthesizer/level generator frequency to 40 MHz.
(15) Press synthesizer/level generator AMPLITUDE key then INCR key.
(16) Press TI MARKER PEAK SEARCH key then [MARKER DELTA] key.
(17) Press synthesizer/level generator INCR key.
(18) Press TI INSTRUMENT STATE SGL SWP key and wait for the completion of
new sweep.
(19) Press TI MARKER MKR key. Adjust TI knob ccw until ' MKR indicates 0 dB
r0.017 dB. The marker should be on the left-hand skirt of the signal. If the marker cannot
be set exactly to 0 dB, note whether the marker is just above or just below the actual -3 dB
point.
(20) Press TI [MARKER DELTA] key. Adjust TI knob cw until ' MKR indicates 0
dB r0.017 dB. The active marker should be on the right-hand skirt of the signal. If the
marker was just above the -3 dB point in (19) above, set the marker just below the -3 dB
point. If the marker was just below the -3 dB point in (19) above, set the marker just above
the -3 dB point.
(21) If TI RES BW setting is 3 kHz or less, record TI ' MKR frequency indication as
corrected -3 dB bandwidth in table 4 and proceed to (24) below.
table 4 for the current TI RES BW setting.
(23) Calculate corrected -3 dB bandwidth using formula below and record the result
in table 4 for the current TI RES BW setting.
Change 1