Table 6. -60 dB Resolution Bandwidth Accuracy
Figure 1. Input attenuator accuracy (50 MHz) - equipment setup.

Calibration Procedure For Spectrum Analyzer Models 8562Ec And 8562Ec-104 Page Navigation
  9    10    11    12    13  14  15    16    17    18    19  
img
. . . . . .
TB 9-6625-2333-24
(40) Press synthesizer/level generator AMPLITUDE key then INCR key.
(41) Press TI MARKER PEAK SEARCH key then [MARKER DELTA] key.
(42) Press synthesizer/level generator INCR ž key.
(43) Press TI INSTRUMENT STATE SGL SWP key and wait for completion of new sweep.
(44) Press TI MARKER MKR key. Adjust TI knob ccw until ' MKR indicates 0 dB
r0.5 dB. The marker should be on the left-hand skirt of the signal. If the marker cannot be
set exactly to 0 dB, note whether the marker is just above or just below the actual -60 dB point.
(45) Press TI [DELTA MARKER] key. Adjust TI knob cw until ' MKR indicates 0
dB r0.5 dB. The active marker should be on the right-hand skirt of the signal. If the
marker was just above the -60 dB point in (44) above, set the marker just below the -60 dB
point. If the marker was just below the -60 dB point in (44) above, set the marker just above
the -60 dB point.
(46) Record TI ' MKR frequency indication as the measured -60 dB bandwidth in
table 6 for the current TI RES BW setting.
(47) Calculate corrected -60 dB bandwidth using formula below and record the result
in table 6 for the current TI RES BW setting.
Corrected -60 dB BW = (Actual Span/Ideal Span) x Measured -60 dB BW
Example: RES BW setting = 1 MHz
Ideal Span = 16 MHz
Actual Span = 17 MHz
Measured -60 dB BW = 9.82 MHz
Corrected -60 dB BW = (17/16) x 9.82 MHz = 10.43 MHz
(48) Calculate selectivity by dividing the corrected -60 dB bandwidth by the corrected
-3 dB bandwidth (from table 4) for the current TI RES BW setting. The result will be <15.
Selectivity = corrected -60 dB BW / corrected -3 dB BW
Example: Selectivity = 10.43 MHz / 0.9415 MHz = 11.08
(49) Press TI keys as listed in (a) through (d) below:
MARKER MKR.
(a)
[MARKERS OFF].
(b)
CONTROL TRIG.
(c)
[SWEEP CONT SGL] to select CONT.
(d)
(50) Repeat technique of (27) (a), (b), and (29) through (49) above for remaining TI
settings in table 5 and table 6.
Change 1
14