Figure 2. Input attenuator accuracy (2.9 GHz) - equipment setup.
Figure 3. IF gain uncertainty - equipment setup.

Calibration Procedure For Spectrum Analyzer Models 8562Ec And 8562Ec-104 Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. . . . . . . . . .
TB 9-6625-2333-24
(27) Press TI keys and enter values using DATA keys as listed in (a) through (d) below:
FREQUENCY then [CENTER FREQ] to 2.9 GHz.
(a)
AMPLITUDE then [REF LVL] to -10 dBm.
(b)
AMPLITUDE then [ATTEN] to 10 dB.
(c)
MARKER MKR [MARKERS OFF].
(d)
(28) Set synthesized signal generator frequency to 2.9 GHz and level output to 0 dBm.
(29) Press TI MARKER MKR key.
(30) Adjust synthesized signal generator level output for a TI MKR indication
between -12.95 and -13.05 dBm.
(31) Press TI keys and enter values using DATA keys as listed in (a) through
(c) below:
(a) MARKER MKR.
(b) [MARKER DELTA].
(c) AMPLITUDE then [ATTEN] to 20 dB.
(32) Wait for completion of sweep. Record ' MKR indication for TI ATTEN 20 dB
setting in table 9.
(33) Repeat technique of (31)(c) and (32) above for remaining TI ATTEN settings
listed in table 9.
(34) Subtract IF GAIN CORRECTION from ' MKR indication and record results
as corrected ' MKR indication for each ATTEN setting listed in table 9. The corrected '
MKR indications will be within limits specified for each setting.
(35) The step-to-step accuracy for TI ATTEN 20 dB setting is equal to the ATTEN
20 dB setting corrected ' MKR indication recorded in table 9.  The corrected ' MKR
indication will be between -0.6 and +0.6 dB.
(36) Calculate the step-to-step accuracy for TI ATTEN 30, 40, 50, 60 and 70 dB
settings by subtracting previous ATTEN setting corrected ' MKR indication recorded in
table 9 from the current ATTEN setting corrected ' MKR indication recorded in table 9.
Difference will be between -0.6 and +0.6 dB for each setting.
EXAMPLE:
(Current)
(Previous)
30 dB step-to-step
TI ATTEN 20 dB
TI ATTEN 30 dB
corrected ' MKR indication 
corrected ' MKR indication
accuracy
=
b. Adjustments. Refer to paragraph 6 c above.
12. IF Gain Uncertainty
a. Performance Check
(1) Connect equipment as shown in figure 3.