TB 9-6625-2333-35
(21) Add -2.27 dB to TI MKR amplitude indication. Calculated displayed average
noise level will be <-148 dBm.
(22) Press TI keys and enter values using DATA keys as listed in (a) through (f) below:
(a)
(b)
[STOP FREQ] to 13.2 GHz.
(c)
MARKER MKR then [MARKERS OFF].
(d)
CONTROL BW then [RES BW] to 1 MHz.
(e)
[VIDEO BW] to 10 kHz.
(f)
[VID AVG ON OFF] to select OFF.
(23) Repeat (17) above.
(24) Add -2.27 dB to TI MKR amplitude indication. Calculated displayed average
noise level will be <-145 dBm.
10. Resolution Bandwidth Accuracy and
Selectivity Test
a. Performance Check
(1) Connect TI 10 MHz REF IN/OUT to synthesizer/level generator 40/N MHz
REF INPUT.
(2) Connect synthesizer/level generator OUTPUT 50 Ω to TI INPUT 50 Ω.
(3) Press synthesizer/level generator keys to values as listed in (a) through (c) below:
(b) AMPLITUDE to -5 dBm.
(c) AMPTD INCR to 1 dB.
(4) Press TI keys enter values using DATA keys as listed in (a) through (k) below:
(a)
INSTRUMENT STATE PRESET.
(b)
INSTRUMENT STATE SAVE.
(c)
[SAVELOCK ON OFF] to select OFF.
(d)
INSTRUMENT STATE CAL.
(e)
[FULL IF ADJ] and wait for IF ADJUST STATUS: message to disappear.
(f)
[IF ADJ ON OFF] to select OFF.
(g)
FREQUENCY then [CENTER FREQ] to 40 MHz.
(h)
SPAN then [SPAN] to 4 MHz.
(i)
AMPLITUDE then [LOG dB/DIV] to 1 dB.
(j)
CONTROL BW then [RES BW] to 2 MHz.
(k)
[VIDEO BW] to 300 Hz.