TB 9-6625-2333-35
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, refer to the manufacturer's manual and perform appropriate
adjustment(s) or corrective action before continuing with the calibration. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. Unless otherwise specified, all controls and control settings refer to TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Connect TI to 115 V ac power source.
b. Press LINE pushbutton to on and allow at least 30 minutes for TI to warm-up and
stabilize.
NOTE
[ ] Denotes softkey on display. Softkeys and keys will be
referred to as keys in this procedure.