TB 9-6625-2333-35
Paragraph
Page
marker accuracy ................................................
Frequency span accuracy.....................................
Frequency response .............................................
Residual responses ...............................................
Final procedure ....................................................
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Spectrum Analyzer, Agilent Models 8562EC and 8562EC-104.
The
manufacturer's manual was used as the prime data source in compiling these instructions.
The equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations. Option 104 deletes the module used to expand user memory
which allows storage and execution of downloadable programs (DLPs) and limit lines.
b. Time and Technique. The time required for this calibration is approximately 6
hours, using the dc and low frequency and microwave technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R)
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications1
Calibrator output accuracy
Amplitude range: -10 dBm
Accuracy: <0.3 dB
See footnotes at end of table.