TB 9-6625-2333-35
Test instrument
MKR
Signal
CENTER FREQ
frequency indications
generator
SPAN
setting
(GHz)
frequency setting
(GHz)
(GHz)
settings
Min
Max
1.5
1.5
10 MHz
1.49948
1.50052
1.5
1.5
20 MHz
1.49895
1.50105
1.5
1.5
50 MHz
1.49745
1.50255
1.5
1.5
100 MHz
1.4948
1.5052
1.5
1.5
1 GHz
1.450
1.550
(8) Press TI keys and enter values using DATA keys as listed in (a) through (f) below:
(a)
FREQUENCY then [CENTER FREQ] to 4 GHz.
(b)
MARKER PEAK SEARCH.
(c)
INSTRUMENT STATE AUX CTRL.
(d)
[INTERNAL MIXER].
(e)
[PRESEL AUTO PK] wait for PEAKING message to disappear.
(f)
SPAN then [SPAN] to 1 MHz.
(9) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
between 3.999988 and 4.000012 GHz.
(10) Press TI SPAN key then [SPAN] key and enter 10 MHz using DATA keys.
(11) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
between 3.99948 and 4.00052 GHz.
(12) Repeat technique of (10) and (11) above for remaining TI SPAN settings listed
Signal generator
Test instrument
CENTER FREQ
MKR frequency indications
frequency
SPAN
setting
(GHz)
setting
(GHz)
(GHz)
settings
Min
Max
4
4
20 MHz
3.99895
4.00105
4
4
50 MHz
3.99745
4.00255
4
4
100 MHz
3.9948
4.0052
4
4
1 GHz
3.950
4.050
(14) Press TI keys and enter values using DATA keys as listed in (a) through (f) below:
(a)
FREQUENCY then [CENTER FREQ] to 9 GHz.
(b)
MARKER PEAK SEARCH.
(c)
INSTRUMENT STATE AUX CTRL.
(d)
[INTERNAL MIXER].
26