TB 9-6625-2333-35
Synthesizer/level
Test instrument
REF LVL
∆MKR indications
generator amplitude
settings
settings
(dB)
(dBm)
(dBm)
Min
Max
-10
-20
-21
-19
-20
-30
-31
-29
-30
-40
-41
-39
-40
-50
-51
-49
-50
-60
-61
-59
-60
-70
-71
-69
-70
-80
-81
-79
a. Performance Check
(1) Connect signal generator RF Output to TI INPUT 50 Ω.
(3) Press TI keys and enter values using DATA keys listed in (a) through (q) below:
(a)
INSTRUMENT STATE PRESET.
(b)
(c)
SPAN then [SPAN] to 1 MHz.
(d)
MARKER PEAK SEARCH.
(e)
MARKER MKR.
(f)
[SIG TRK ON OFF] to select ON.
(g)
SPAN then [SPAN] to 5 kHz wait for signal to be centered in a 5 kHz span.
(h)
CONTROL BW then [RES BW] to 10 Hz.
(i)
SPAN then [SPAN] to 100 Hz wait for signal to be centered in a 100 Hz span.
(j)
MARKER MKR.
(k)
[SIG TRK ON OFF] to select OFF.
(l)
AMPLITUDE then [LOG dB/DIV] to 2 dB.
MARKER PEAK SEARCH.
(m)
MARKER MKR→.
(n)
[MARKER→ REF LVL].
(o)
(p)
MARKER PEAK SEARCH.
(q)
[MARKER DELTA]
(4) Adjust TI knob ccw for a ∆ MKR indication between -9.5 and -10.5 dB.
(5) Press TI keys as listed in (a) through (d) below:
(a) [MARKER NORMAL].
(b) [MARKER → CF].