TB 9-6625-2335-24
(4) Enter .003,3457 using MATH keys.
(5) Enter .03,3457 using MATH keys.
(6) Enter .3,3457 using MATH keys.
(7) Enter 1,3457 using MATH keys.
(8) Press ENT key. TI will display CALIBRATING while adjustment routine
is running.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI INPUT I and LO.
(2) Press ACI key.
(3) Press RANGE or key to select ranges and set calibrator for output settings
at 1 kHz listed in table 13. If TI does not indicate within limits specified, perform
corresponding software adjustments.
Test instrument
indications
Test instrument
Calibrator output
Software
ranges
settings at 1kHz
Adjustments
Min
Max
b (1), (2), and (5)
30 mA
30 mA
29.873
mA
30.127
mA
b (1), (3), and (5)
300 mA
300 mA
298.73
mA
301.27
mA
b (1), (4), and (5)
1A
1A
0.9929
A
1.0071
A
b. Software Adjustments
(1) Press blue shift key then CONFIGURATION C/OFFSET COMP key. TI will
display CAL.
(2) Enter .03,3457 using MATH keys.
(3) Enter .3,3457 using MATH keys.
(4) Enter 1,3457 using MATH keys.
(5) Press ENT key. TI will display CALIBRATION while adjustment routine
is running.
a. Performance Check
f
(1) Connect function/arbitrary generator Output to TI INPUT HI and LO using 50
eedthrough termination.
(2) Press FREQ key.
(3) Set function/arbitrary generator for a sine wave, 20 Hz, 1 V rms output. If TI
does not indicate between 19.99000 and 20.01000 Hz, perform b below.
(4) Set function/arbitrary generator frequency to 1 MHz. If TI does not indicate
between 999.9000 kHz and 1.000100 MHz, perform b below.