Calibration Procedure For Multimeter, Model 3457A Page Navigation
  7    8    9    10    11  12
TB 9-6625-2335-24
b. Software Adjustments
(1) Set function/arbitrary generator frequency to 1 MHz.
(2) Press blue shift key then CONFIGURATION C/OFFSET COMP key. TI will
display CAL.
(3) Enter 1000000,3457 using MATH keys.
(4) Press ENT key. TI will display CALIBRATING while the adjustment routine
is running.
15. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.