TB 9-6625-2337-24
Table 31. Azimuth Gate Characteristics - Continued
TI
Oscilloscope
Test description
Settings
Measurement type
Minimum
Maximum
AZIMUTH
----------------------------------------
Width of pulse
250.2 ms
305.8 ms
train on channel 2
1
AZIMUTH GATE
Delta time
1.350 s
1.650 s
Press FUNC,
1500 ms
Channel 1 to
4, ENTR (PRF/PRI)
channel 2 first
Use the arrow keys to highlight
pulses
PRF:
300, ENTR (300 PPS)
9
Press FUNC,
, ENTR (ACTIVE GATING)
Use the arrow keys to highlight
Ant. Rate:
4000000, ENTR (4,000,000 s)
Use the arrow keys to highlight
AZ GATE START:
135, ENTR (135 )
Use the arrow keys to highlight
AZ GATE WIDTH:
3, ENTR (3.0 )
AZIMUTH
----------------------------------------
Width of pulse
29.7 ms
36.3 ms
WIDTH
train on channel 2
33 ms
(7) Move connection from RADAR SIM POS READOUT GATE to TRIGGERS
0 OUT.
(8) Move connection from RADAR SIM NORTH TRIGGER OUT to RADAR SIM
POS READOUT GATE.
(9) Adjust the oscilloscope to trigger on channel 1 with channel 2 pulse displayed to
the left of the pulse on channel 1. Set oscilloscope to measurement type to delta time.
(10) Oscilloscope will indicate within limits listed in first row of table 32.
(11) Repeat technique of (3) through (5) above using remaining setting and limits
TI
Oscilloscope
Test description
Connections
Measurement type
Minimum
Maximum
AZIMUTH GATE
----------------------------------------
Delta time
270 μs
330 μs
300 ms
Channel 2 to
channel 1 pulse
POS READOUT
----------------------------------------
Width of pulse on
270 μs
330 μs
PW
channel 1
POS READOUT
----------------------------------------
V Top
15 V
25 V
AMP
AZIMUTH GATE
Move the connection from the
Delta time
270 μs
330 μs
RADAR SIM POS READOUT
Channel 2 to
GATE to the RADAR SIM NEG
channel 1 pulse
READOUT GATE.
43