TB 9-6625-2337-35
TI
Oscilloscope
Channel 1
Period
measurement
Test description
Trigger setting
Amp
(s)
Width
type
Minimum
Maximum
143mS EXT TRIG
0 TRIGGGER:
2
143 m
300 n
6.5 Hz
7.5
EXTERN +
10k EXT TRIG
0 TRIGGGER:
-15
100 u
10 u
9.9995 kHz
10.0005
EXTERN -
kHz
0 TRIGGER:
-------
-------
-------
Max
15 V
25 V
0 OUT 20 V level1
INT
1 s
2 s
0 OUT 1.5 s PW
0 TRIGGER:
-------
-------
-------
Width
INT
0 OUT Overshoot
0 TRIGGER:
-------
-------
-------
Overshoot
.01 %
5%
INT
0 OUT Rise Time
0 TRIGGER:
-------
-------
-------
Rise Time
200 ns
INT
0 OUT Decay Time
0 TRIGGER:
-------
-------
-------
Fall Time
200 ns
INT
1 s
2 s
TTL 1.5 s width
0 TRIGGER:
-------
-------
-------
Channel 2
INT
Width
TTL 5V level
0 TRIGGER:
-------
-------
-------
Channel 2 Top
2.5 V
5.5 V
INT
1
(12) Verify that the pulse displayed on channel 2 is approximately coincident with
the pulse on channel 1.
(13) Disconnect the equipment setup.
b. Adjustments. No further adjustments can be made.
a. Performance Check
(1) Connect TI TRIGGERS 0 OUT to the oscilloscope channel 1 input using a 75 Ω
feedthrough termination.
(2) Connect TI VAR PULSE OUTPUT to the oscilloscope channel 2 input using a
75 Ω feedthrough termination.
(3) Press TI keys as listed in (a) through (g) below:
(a)
Press FUNC and ENTR keys.
(b)
Use the arrow keys to highlight TRANSPONDER MENUS and press
ENTR key.
(c)
6, ENTR (VAR PULSE).
(d)
Use the arrow keys to highlight VP1 PW:.
(e)
.25, ENTR.
(f)
Use the arrow keys to highlight VP1 DLY:.
(g)
2, ENTR.