TB 9-6625-2338-24
(18) Press and hold down the TI BSP key then press and release the TI OUTPUT
key until TI display indicates Offset B
###.
(19) Release the TI BSP key.
(20) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the V top measured value displayed on the oscilloscope is equal to
0.0 mV r 10 mV (R).
(21) Disconnect oscilloscope Vertical 1 input from B output.
(22) Connect oscilloscope Vertical 1 input to +AB output.
(23) Press and hold down the TI BSP key then press and release the TI OUTPUT
key until TI display indicates Offset AB
###.
(24) Release the TI BSP key.
(25) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the V top measured value displayed on the oscilloscope is equal to
0.0 mV r 10 mV (R).
(26) Disconnect oscilloscope Vertical 1 input from +AB output.
(27) Connect oscilloscope Vertical 1 input to AB output.
(28) Press and hold down the TI BSP key then press and release the TI OUTPUT
key until TI display indicates Offset AB
###.
(29) Release the TI BSP key.
(30) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the V top measured value displayed on the oscilloscope is equal to
0.0 mV r 10 mV (R).
(31) Disconnect oscilloscope Vertical 1 input from AB output.
(32) Connect oscilloscope Vertical 1 input to C output.
(33) Press and hold down the TI BSP key then press and release the TI OUTPUT
key until TI display indicates Offset C
###.
(34) Release the TI BSP key.
(35) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the V top measured value displayed on the oscilloscope is equal to
0.0 mV r 10 mV (R).
(36) Disconnect oscilloscope Vertical 1 input from C output.
(37) Connect oscilloscope Vertical 1 input to D output.
(38) Press and hold down the TI BSP key then press and release the TI OUTPUT
key until TI display indicates Offset D
###.
(39) Release the TI BSP key.
(40) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the V top measured value displayed on the oscilloscope is equal to
0.0 mV r 10 mV (R).
(41) Disconnect oscilloscope Vertical 1 input from D output.
9