Output Amplitude
Table 5. Output Offset

Calibration Procedure For Digital Delay/Pulse Generator Dg535 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . .
TB 9-6625-2338-24
(20) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the oscilloscope waveform amplitude is 0.800 Vpp (R).
(21) Disconnect oscilloscope Vertical 1 input from TI +AB output.
(22) Connect oscilloscope Vertical 1 input to TI C output.
(23) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude C
###.
(24) Release the TI BSP key.
(25) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the oscilloscope waveform amplitude is 0.800 Vpp (R).
(26) Disconnect oscilloscope Vertical 1 input from TI C output.
(27) Connect oscilloscope Vertical 1 input to TI D output.
(28) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude D
###.
(29) Release the TI BSP key.
(30) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the oscilloscope waveform amplitude is 0.800 Vpp (R).
(31) Disconnect oscilloscope Vertical 1 input from TI D output.
(32) Connect oscilloscope Vertical 1 input to TI +CD output.
(33) Press and hold down the TI BSP key then press and release the TI GPIB key
until TI display indicates Amplitude +CD
###.
(34) Release the TI BSP key.
(35) Use the TI left/right (4, 6) and the up/down (8, 2) cursor control keys to adjust
the displayed value until the oscilloscope waveform amplitude is 0.800 Vpp (R).
10. Output Offset
a. Performance Check
(1) Connect TI T0 output to oscilloscope Vertical 1 input.
(2) Press TI keys as listed in (a) through (f) below:
Press RECALL, 0, EXC.
(a)
4 (cursor left), 4.
(b)
TRIG, NUM/REM, 100000, EXC.
(c)
DELAY, DELAY, NUM/REM, 5, EXP, , 6 (cursor right), EXC.
(d)
DELAY, DELAY, NUM/REM, 5, EXP, , 6, EXC.
(e)
OUTPUT, OUTPUT, OUTPUT, 6, EXC.
(f)
(3) Set oscilloscope for a display of 4 vertical divisions, 4 pulses and to measure V
top. Measure TI T0 output. If oscilloscope does not indicate within limits shown in first
row of table 5, perform b below.
(4) Move oscilloscope Vertical 1 input connection from TI T0 output to TI A output.