Equipment Setup
Output Amplitude

Calibration Procedure For Digital Delay/Pulse Generator Dg535 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . .
TB 9-6625-2338-24
Press RECALL, 0, EXC.
(a)
4 (cursor left), 4.
(b)
TRIG, NUM/REM, 100000, EXC.
(c)
DELAY, DELAY, NUM/REM, 5, EXP, , 6 (cursor right), EXC.
(d)
DELAY, DELAY, NUM/REM, 5, EXP, , 6, EXC.
(e)
OUTPUT, OUTPUT, OUTPUT, 6, EXC.
(f)
(3) Configure oscilloscope for a peak to peak measurement and measure TI T0 NIM
output. If oscilloscope does not indicate within limits shown in T0/NIM row of table 4,
perform b below.
Table 4. Output Amplitude
Test instrument
Oscilloscope indications
Output
Mode
Min
Max
T0
NIM
726 mV
874 mV
T0
TTL
3.83  V
4.17  V
T0
VAR
2.86  V
3.14  V
T0
ECL
-1.904 V
-1.696 V
A
NIM
726 mV
874 mV
A
TTL
3.83  V
4.17  V
A
VAR
2.86  V
3.14  V
A
ECL
-1.904 V
-1.696 V
B
NIM
726 mV
874 mV
B
TTL
3.83  V
4.17  V
B
VAR
2.86  V
3.14  V
B
ECL
-1.904 V
-1.696 V
+AB
NIM
726 mV
874 mV
+AB
TTL
3.83  V
4.17  V
+AB
VAR
2.86  V
3.14  V
+AB
ECL
1.904 V
1.696 V
C
NIM
726 mV
874 mV
C
TTL
3.83  V
4.17  V
C
VAR
2.86  V
3.14  V
C
ECL
1.904 V
1.696 V
D
NIM
726 mV
874 mV
D
TTL
3.83  V
4.17  V
D
VAR
2.86  V
3.14  V
D
ECL
1.904 V
1.696 V
+CD
NIM
726 mV
874 mV
+CD
TTL
3.83  V
4.17  V
+CD
VAR
2.86  V
3.14  V
+CD
ECL
1.904 V
1.696 V
(4) Select TI T0 output TTL mode by pressing 4 and measure output. If oscilloscope
does not indicate within limits shown in T0/TTL row of table 4, perform b below.
(5) Press TI key 6, 6, 6.
(6) Press TI OUTPUT key and set output amplitude to 3.00 V.
(7) Press TI OUTPUT key and set output offset to 0.00 V.