TB 9-6625-2339-35
Test instrument parameters
Performance specifications
10 MHz Reference Accuracy
Frequency: 10 MHz
Accuracy: 0.01 ppm
Range: 100 Hz
and selectivity
Accuracy: <30%
Range: 300 Hz to 300 kHz
Accuracy: <10%
Range: 1 MHz and 2 MHz
Accuracy: <25%
Selectivity: 60 dB/3 dB bandwidth ratio: <15:1
Input attenuator accuracy
Range: 0 to 60 dB, referenced to 10 dB input attenuation
Frequency range: 10 kHz to 22 GHz1
Accuracy: <1.8 dB/10 dB step, 3.5 dB maximum
IF gain uncertainty
Range: 0 to -80 dBm reference level, 10 dB input attenuation
Accuracy: <1.0 dB
Residual FM (zero span)
<5 Hz X N p-p in 20 mS
Noise sidebands
Range: 1.0 GHz, 0 dBm 10, 20, 30 and 100 kHz offset
Accuracy: <-90 dBc @ 10 kHz offset
<-98 dBc @ 20 kHz offset
<-100 dBc @ 30 kHz offset
<-112 dBc @ 100 kHz offset
Frequency Readout Accuracy
Frequency range: 9 kHz to 26.5 GHz1
Accuracy: <[ 0.5% of span +15% of resolution bandwidth + 50 Hz
Marker Frequency Count
Accuracy
Accuracy: <[ (marker frequency x reference frequency error ) +
(50 Hz x N) + LSD]
Frequency Span Readout
Range: 100 Hz to 1 GHz in a 1,2,5, sequence
Accuracy
Accuracy <5% of the selected span over the center 80% of the
graticule area
Frequency response (10 dB
input attenuation)
Displayed average noise level
Frequency Range
(dBm)
(no signal at input, 100 Hz
RES BW, and 0 dB input
10 kHz
<-90
attenuation)
100 kHz
<-100
1 MHz to 2.9 GHz
<-121
2.9 to 6.46 GHz
<-121
6.46 to 13 GHz
<-110
13 to 19.7 GHz1
<-105
19.7 to 22 GHz1
<-100
Residual responses (no signal
Range: 200 kHz to 6.46 GHz
at input, 0 dB input
Responses: <-90 dBm
attenuation)
Frequency drift
<50 Hz X N2 per minute of sweep time spans <100 kHz
<2 kHz X N2 per minute of sweep time spans 101 kHz to <1 MHz
See footnotes at end of table.