TB 9-6625-2339-35
Table 2. Minimum Specifications of Equipment Required - Continued
Common name
Minimum use specifications
Manufacturer and model
(part number)
SYNTHESIZER/LEVEL
Hewlett-Packard, Model
Amplitude range: -80 to +12 dBm
3335AOPT001-KO6 (MIS-35938)
Accuracy: 0.02 dB per 10 dB step
TIMER/COUNTER
Fluke, Model PM6681
ANALYZER
Resolution: 0.001 Hz
(PM6681/656)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in TM 43-6625-914-40.
d. When indications specified in
paragraphs 8 through 19 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 19. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to TI.
HIGH VOLTAGE is used or exposed during the performance of
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
NOTE
The analyzer must sit with the power off for at least 60
minutes before performing the following test. This adequately
simulates a cold start.
NOTE
Throughout this procedure soft keys are identified by the use of
brackets [ ] around the key.