TB 9-6625-2340-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
1
>+5 dBm: 2.3 GHz to 6.1 GHz
100 Hz to 0.7 GHz: <-40 dBm <-80 dBc
.07 to 2.5 GHz: <-40 dBm <-80 dBc
2.0 to 18 GHz: <-10 dBm <100 dBc
Intermodulation distortion
Third order intercept:
>+5 dBm: 100 Hz to 5 MHz
>+7 dBm: 5 MHz to 5.8 GHz
>_5 dBm: 5.8 GHz to 18 GHz
Image and out of band response
<-70 dBc: 100 Hz to 18 GHz
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287, AN/GSM-705 or Secondary Reference Set NSN 4931-00-621-
7878. Alternate items may be used by the calibrating activity. The items selected must be
verified to perform satisfactorily prior to use and must bear evidence of current calibration.
The equipment must meet or exceed the minimum use specifications listed in table 2. The
accuracies listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
50 , 20 W dummy load.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
ATTENUATOR NO. 1
Range: 0 to 12 dB
Hewlett-Packard, Model 355C (7910807)
Frequency: 50 MHz
ATTENUATOR NO 2
Range: 0 to 60 dB
Hewlett-Packard, Model 355D (355D)
Frequency: 300 MHz
ELECTRONIC COUNTER
Range: 20 ms to 200 s
Fluke, Model PM6681/656
Accuracy: 10%
(PM6681/656)
Frequency range:
(SG-1288/G)
2 kHz to 100 kHz triangle wave
@ 1 V amplitude
FUNCTION/ARBITRARY
Agilent, Model 33250A (33250A)
Sweep time: 1 sec to 99.9 sec
LOW PASS FILTER1
fc 250 MHz
K&L Model 5L380-250-B/B (5L380-250-
B/B)
LOW PASS FILTER1
fc 1200 MHz
Hewlett-Packard, Model 360B (360B)
LOW PASS FILTER1
fc 8 GHz
K&L Model 6L250-8000-NP/N (6L250-
8000-NP/N)
Measuring receiver system N5531S
MEASURING RECEIVER
Frequency range: 300 MHz
consisting of: Spectrum Analyzer Agilent,
Accuracy: 300 Hz
Model E4440A (E4440A), Power Meter
dBm range:-9.7 to -10.3
Agilent, Model E4419B (E4419B), and
Accuracy: 0.075 dBm
Sensor Module Agilent, Model 518 (518)
See footnote at end of table.