Figure 2. Calibrator amplitude adjustment.
Table 4. Center Frequency Accuracy - Continued

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. . . . . .
TB 9-6625-2340-24
10. Center Frequency Accuracy
a. Performance Check
(1) Connect TI CAL OUTPUT to TI RF INPUT.
(2) Press 2 22 GHz, RECALL, 9 keys.
(3) Adjust FREQ ZERO control for a maximum amplitude trace.
(4) Press 2 22 GHz key.
(5) Disconnect TI CAL OUTPUT from TI RF INPUT.
(6) Connect TI RF INPUT to synthesized signal generator No. 1 RF Output.
(7) Connect TI 10 MHz OUT (rear panel) to the synthesized signal generator No. 1
10 MHz REF IN (rear panel).
(8) Set synthesized signal generator No. 1 for an output of 2.000000 GHz at a level
of 0 dBm.
(9) Press TI keys as listed in (a) through (d) below:
CENTER FREQUENCY, 2, GHz.
(a)
FREQUENCY SPAN, 1, MHz.
(b)
MARKER PEAK SEARCH.
(c)
MKR
CF.
(d)
(10) Press TI REFERENCE LEVEL and adjust as necessary to place signal peak at
convenient level.
(11) If TI center frequency readout does not fall within limits listed in the first row of
table 4, perform b below.
(12) Repeat technique of (8) through (10) above for remaining center frequencies and
frequency spans listed in table 4. If TI center frequency readout does not indicate within
limits specified in table 4, perform b below.
Table 4. Center Frequency Accuracy
Synthesized
Test instrument
signal generator
Frequency readout
Output
Center
frequency
frequency
Frequency
Min
Max
(GHz)
(GHz)
span
(GHz)
(GHz)
2
2
1  MHz
1.99998
2.00002
2
2
10  MHz
1.9997
2.0003
2
2
100  MHz
1.998
2.002
2
2
1  GHz
1.98
2.02
3
3
1  MHz
2.99998
3.00002
3
3
10
MHz
2.9997
3.0003
3
3
100  MHz
2.998
3.002
3
3
1  GHz
2.98
3.02
6
6
1  MHz
5.99998
6.00002
8