Figure 33. A8 board.

Calibration Procedure For Spectrum Analyzer, Model 8566B Page Navigation
  77    78    79    80    81  82
img
. . . . . . . .
TB 9-6625-2340-24
(31) Multimeter should indicate 5.2 0.05 V dc.
(32) Connect multimeter HI to A18TP5 (fig. 34) and multimeter LO to TI chassis.
(33) The 5.2 V indicator A18DS3 (fig. 34) (yellow LED) should be lit.
(34) Multimeter should indicate 5.2 0.05 V dc.
(35) Connect multimeter HI to A18TP1 (fig. 34) and multimeter LO to TI chassis.
(36) The 40 V indicator A18DS1 (fig. 34) (yellow LED) should be lit.
(37) Multimeter should indicate 39.8 0.4 V dc.
(38) Connect multimeter HI to A18TP4 (fig. 34) and multimeter LO to TI chassis.
(39) The 10 V indicator A18DS2 (fig. 34) (yellow LED) should be lit.
(40) Multimeter should indicate 10.0 0.1 V dc.
29. Final Procedure
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.