TB 9-6625-2340-35
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Frequency range: 300 MHz
Consisting of: Measuring Receiver
Accuracy: 300 Hz
Hewlett-Packard, Model 8902A
dBm range:-9.7 to -10.3
(8902A), Microwave Converter
Accuracy: 0.075 dBm
Hewlett-Packard Model 11793A
(11793A), and Sensor Modules,
Hewlett-Packard, Model 11722A
(11722A) and 11792A (11792A)
MICROWAVE
Frequency range: 2 GHz to 18 GHz
Hewlett-Packard, Model 5352BOPT001
FREQUENCY COUNTER
Accuracy: 1 kHz 1 count
(5352BOPT001)
Range: 1000 Vdc
John Fluke, Model 8840A/AF
Accuracy: 2 Vdc
(AN/GSM 64D)
Range: Vert. 2 Volts / cm
(OS-303/G)
Horiz. 100 ms / cm
POWER SPLITTER
Range: 20 MHz to 18 GHz
Weinschel, Model 1870 (1870)
-10
TIME/FREQUENCY
Resolution: 1 x 10
Datum, Model ET6000-75
(13589305)
WORKSTATION
(SG-1207/U)
NO. 1
Power range: 0 to -15 dBm
Accuracy: 2 dBm
(SG-1219/U) or Anritsu, Model 68369NV
Frequency range: 2 GHz to 18 GHz
(68369NV)
NO. 2
Power range: 0 to -120 dBm
Accuracy:1.525 dBm
(SG-1219/U) or Anritsu, Model 68369NV
Frequency range: 2 GHz to 18 GHz
(68369NV)
NO. 3
Power range:
0 to -120 dBm
Accuracy:
1.525 dBm
SYNTHESIZER/FUNCTION
Agilent, Model 33250A (33250A)
GENERATOR
Sweep time: 1 sec to 99.9 sec
Hewlett-Packard, Model 3335AOPT001-
SYNTHESIZER/LEVEL
Frequency range: 10 kHz to 50 MHz
KO6 (MIS-35938)
GENERATOR
Amplitude range: -80 to +10 dBm
Accuracy:
0.02 dB per 10 dB
step
1Limited
deploy.
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in the manufacturer's manuals.
d. When indications specified in
paragraphs 8 through 27 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are
CHANGE 2