TB 9-6625-2340-35
(59) Multimeter should indicate between 8.5 and 11.0 V dc. If measurement is out of
tolerance, adjust A4A5R2 for a multimeter indication of 9.5 0.1 V dc (R).
(60) Connect equipment as shown in figure 16 below.
(61) Set both attenuators to 0 dB.
(62) Disconnect 97 (white/violet) cable from A4A8J1 (fig. 13).
(63) Disconnect 89 (gray/white) cable from A4A6J1 (fig. 9).
(64) Connect 97 (white/violet) cable to A4A6J1 (fig. 9).
(65) Press TI keys as listed in (a) through (e) below:
(a)
RES BW, 1, MHz.
(b)