TB 9-6625-2340-35
(f) SCALE LIN.
(g) MARKER ∆.
(h) RES BW, 1, MHz.
(5) Adjust A4A6R29 (fig. 9) to align markers on TI display. MKR ∆ level should
indicate 1.00X.
(6) Press TI keys as listed in (a) and (b) below:
(a) 2-22 GHz.
(b) RECALL, 9.
(7) Adjust TI FREQ ZERO for maximum trace amplitude.
NOTE
The pictured filter (fig. 10) is required to continue with this
adjustment procedure.
(8) Press TI keys as listed in (a) through (e) below:
25