TB 9-6625-2340-35
Test
Electronic counter
instrument
displayed reading
Sweep time
Min
Max
20 s
18 s
22 s
200 s
188 s
222 s
b. Adjustments. None
19. Line
Related Sidebands
a. Performance Check
(1) Connect TI CAL OUTPUT to TI RF INPUT.
(2) Press TI 2-22 GHz, RECALL, 9.
(3) Adjust TI FREQ ZERO for a maximum amplitude trace.
(4) Press TI keys as listed in (a) through (c) below:
(a)
2-22 GHz.
(b)
CENTER FREQUENCY, 100, MHz.
(c)
(5) Wait for asterisk in upper right corner of TI display to disappear.
(6) Press TI keys as listed in (a) through (c) below:
(a)
MARKER PEAK SEARCH.
MKR→CF.
(b)
MKR→REF LVL.
(c)
(7) Wait for asterisk in upper right corner of TI display to disappear. Trace should
be centered on display.
(8) Press TI keys as listed in (a) and (b) below:
(a) SHIFT, VIDEO BWG.
(b) SWEEP SINGLE, 10, Hz.
(9) When TI display VID AVG readout is 10, press TI keys as listed in (a) through
(e) below:
(a)
SHIFT.
(b)
TRACE B BLANK.
(c)
TRACE A VIEW.
(d)
MARKER PEAK SEARCH.
MARKER ∆.
(e)
(10) Adjust TI DATA knob to move the ∆ MARKER to the point on the displayed
trace at which the ∆ marker frequency indication is -120 Hz.
(11) TI marker ∆ amplitude will indicate within limits specified in first row of table
58