TB 9-6625-2340-35
(7) Wait until TI display VID AVG readout reaches 15, and then press TI keys as
listed in (a) through (c) below:
(a)
SHIFT.
(b)
TRACE B BLANK.
(c)
MARKER NORMAL.
(8) Tune TI to 100 Hz reading on TI display.
(9) Read the amplitude indicated on TI display. Amplitude indicated will be within
Test instrument
Center
Displayed amplitude
limit (dBm)
100 Hz
< -95
(10) Press TI keys as listed in (a) through (c) below:
(a)
CENTER FREQUENCY, 51, kHz.
(b)
FREQUENCY SPAN, 0, Hz.
(c)
SWEEP TIME, 20, mSEC.
(11) Wait until TI display VID AVG readout reaches >20.
(12) Read the amplitude indicated on TI display. Amplitude indicated will be within
limits specified in the first row of table 26.
(13) Repeat technique of (10)(a), (11) and (12) above for remaining CENTER
FREQUENCY settings in table 26. TI
amplitude displayed will
indicate within
limits
Test instrument
Center
Displayed amplitude
limit (dBm)
51
kHz
< -112
2.0
MHz
< -134
1.0
GHz
< -134
< -134
2.499
GHz
2.510
GHz
< -132
5.799
GHz
< -132
< -125
5.810
GHz
12.499
GHz
< -125
< -119
12.510
GHz
18
GHz
< -119
b. Adjustments. None