TB 9-6625-2343-50
(24) Set calibrator for a wideband output frequency of 10 kHz at a level of 179.4 mV.
(25) Adjust calibrator level for a TI indication of 25 kHz.
(26) Press TI MEASUREMENT ΦM key.
(27) If TI indication is not within limits specified in the first row of table 10, perform
b below.
(28) Repeat technique of (23)(c) and (24) through (26) above for remaining calibrator
initial settings and TI FM deviations in table 10. If TI indications are not within limits
Test instrument
Φ Modulation indications
Calibrator
FM deviation
initial output
indication
(rad)
Frequency
Level (mV)
(kHz)
Max
Max
10 kHz
179.4
25
2.39
2.61
10 kHz
18
2.5
0.239
0.261
1 kHz
178.5
25
24.0
26.0
200 Hz
84
5
24.0
26.0
(29) Move TI sensor module from AM/FM test source FM/4 OUTPUT to FM OUTPUT.
(30) Press TI keys as listed in (a) through (e) below:
(a)
AUTOMATIC OPERATION.
MEASUREMENT ΦM.
(b)
(c)
DETECTOR PEAK+.
(d)
HP FILTER 300 Hz.
(e)
LP FILTER 15 kHz.
(31) Set audio analyzer for an output of 1 kHz source frequency, 1 mV level step, 600 Ω
output impedance, 4.38 V output level, and activate HP filter.
(32) Disconnect calibrator from AM/FM test source.
(33) Connect audio analyzer OUTPUT HIGH to AM/FM test source AUDIO INPUT.
(34) Adjust audio analyzer source level for a TI indication of 400 rad.
(35) Enable audio analyzer special function 15, and then activate distortion function.
(36) If audio analyzer indication is not ≤ 0.1%, perform b below.
(37) Set audio analyzer for an output level of 440 mV and 1 mV level steps.
(38) Adjust audio analyzer source level for a TI indication of 40 rad.
(39) Press TI keys as listed in (a) through (c) below:
(a) DATA 3.1.
(b) DATA SPCL.
(c) MEASUREMENT ΦM.
(40) Enable audio analyzer special function 15, and then activate distortion function.
(41) If audio analyzer indication is not ≤ 0.1%, perform b below.
(42) Reduce all outputs to minimum and disconnect equipment setup.
25