TB 9-6625-2343-50
(8) Using the formula below, calculate TI ref power. If calculated power is not
between 0.988 and 1.012, perform b below.
5 X ((V off2 - V on2) C) = Ref power
(9) Disconnect equipment setup.
b. Adjustments
(1) Disconnect equipment setup.
(2) Connect TI CALIBRATION RF POWER OUTPUT to TI sensor module.
(3) Press TI keys as listed in (a) through (e) below:
(a)
BLUE.
(b)
INSTR PRESET.
(c)
DATA 45.16.
(d)
DATA SPCL.
(e)
TRACK MODE.
(4) If TI frequency indication is not between 49.5 and 50.5 MHz remove the thumb
screw retaining the A32 oscillator and slide the assembly out. Adjust A32L1 (fig. 17)
for 50 MHz 0.5 MHz.
R4 LEVEL
A32
ADJ
POWER
POWER REFERENCE
ADJUST
(5) Reinstall A32 assembly.
NOTE
The TEGAM IIA measuring system must be warmed up for at
least 2 hours and connected in standard configuration prior to
taking measurements.
(7) Press multimeter keys as listed in (a) through (e) below:
(a) Blue shift, RESET.
34