TB 9-6625-2345-35
b. Adjustments.
(1) Disconnect equipment setup.
(2) Press and release TI spectrum analyzer POWER key to OFF.
(3) Disconnect TI power cord.
(4) Set TI spectrum analyzer on its front panel.
(5) Loosen four screws through the back rubber feet.
(6) Pull cover up and off.
(7) Place TI on its top and reconnect power cord.
(8) Connect TI CAL OUT to electronic counter input C.
(9) Depress TI spectrum analyzer POWER key to ON.
(10) Remove the access screw from A1A37 (fig. 1) and adjust oscillator for a
100,000,000 Hz indication on electronic counter.
(11) Disconnect TI CAL OUT from electronic counter INPUT 2.
(12) Press and release TI spectrum analyzer POWER key to OFF.
(13) Disconnect power cord.
(14) Reinstall access screw on A1A37 module.
(15) Set TI spectrum analyzer on its front panel.
(16) Replace TI outer cover.
(17) Tighten four screws through back rubber feet.
(18) Place TI in its normal orientation and reconnect power cord.
(19) Depress TI spectrum analyzer POWER key to ON.
6