dBm.
(14) Set measuring receiver to measure RF POWER in dBm at 90 MHz.
(15) Adjust signal generator #1 output amplitude for a measuring receiver of
10.0 0.1 dBm.
(16) Press TI keys as listed in (a) through (e) below:
(a) BLUE SHIFT.
(b) FREQ, 9, 0, MHz +dBX.
(c) SINGLE SWEEP.
(d) GREEN SHIFT.
(e) PEAK FIND.
(17) TI MKR LEVEL indication will be within limits specified in first row of table
(18) Repeat technique of (13) and (16) above for remaining signal generator #1
frequencies and test instrument settings in table 11. TI MKR LEVEL indications will be
24