TB 9-6625-2345-35
(6) Press TI keys as listed in (a) through (e) below:
(a) BLUE SHIFT.
(b) FREQ, 2, 0, 0, kHz dBX.
(c) SINGLE SWEEP.
(d) GREEN SHIFT.
(e) PEAK FIND.
(7) TI MKR LEVEL indication will be within limits specified in first row of table
(8) Repeat technique of (5) and (6) above for remaining synthesizer/level generator
frequencies and test instrument settings in table 10. TI MKR LEVEL indications will be
Table 10. Frequency Response
200 kHz to 80 MHz
Synthesizer/level
Test instrument
generator
Frequency
Mkr level indication
(dB)
Min
Max
200 kHz
200 kHz
1.0
1.0
300 kHz
300 kHz
1.0
1.0
400 kHz
400 kHz
1.0
1.0
500 kHz
500 kHz
1.0
1.0
600 kHz
600 kHz
1.0
1.0
700 kHz
700 kHz
1.0
1.0
800 kHz
800 kHz
1.0
1.0
900 kHz
900 kHz
1.0
1.0
1 MHz
1 MHz
1.0
1.0
10 MHz
10 MHz
1.0
1.0
20 MHz
20 MHz
1.0
1.0
30 MHz
30 MHz
1.0
1.0
40 MHz
40 MHz
1.0
1.0
50 MHz
50 MHz
1.0
1.0
60 MHz
60 MHz
1.0
1.0
70 MHz
70 MHz
1.0
1.0
80 MHz
80 MHz
1.0
1.0
(9) Reduce synthesizer/level generator output to minimum and disconnect
synthesizer/level generator OUTPUT 50Ω from TI RF INPUT 50Ω.
(10) Connect 11722A sensor module to measuring receiver RF POWER OUTPUT.
(11) Zero and calibrate sensor module.
(12) Connect equipment as shown in figure 8 below.
23