TB 9-6625-2345-35
(10) Set TIME/DIV control to 10 ms.
(12) Adjust A1A72R1062 (fig. 7) for one marker per division (R).
(13) Reduce all outputs to minimum and disconnect equipment setup.
(14) Press and release TI spectrum analyzer POWER key to OFF.
(15) Disconnect TI power cord.
(16) Set TI spectrum analyzer on front panel.
(17) Slide outer cover onto TI.
(18) Tighten four screws through back rubber feet.
(19) Place TI in its normal orientation and reconnect power cord.
(20) Depress TI spectrum analyzer POWER key to ON.
15. Resolution BW and
Shape Factor
a. Performance Check
(1) Connect TI CAL OUT to TI RF INPUT 50Ω.
(2) Press TI keys as listed in (a) through (j) below:
(a)
BLUE SHIFT.
(b)
RESET.
(c)
BLUE SHIFT.
(d)
FREQ, 1, 0, 0, MHz +dBX.
(e)
BLUE SHIFT.
(f)
REF LEVEL, 2, 0, kHz dBX.
(g)
BLUE SHIFT.