TB 9-6625-2345-35
(3) Set TI spectrum analyzer on front panel.
(4) Loosen four screws through back rubber feet.
(5) Pull cover up and off.
(6) Place TI on bottom and reconnect power cord.
(7) Depress TI spectrum analyzer POWER key to ON.
(8) Connect pulse generator OUTPUT to TI RF INPUT 50Ω.
(9) Press TI keys as listed in (a) through (h) below:
(a) BLUE SHIFT.
(b) RESET.
(c) BLUE SHIFT.
(d) FREQ, 6, 0, 0, MHz +dBX.
(e) BLUE SHIFT.
(f) SPAN/DIV, 1, 0, 0, MHz +dBX.
(g) BLUE SHIFT.
(h) REF LEVEL, 0, kHz dBX.
(10) Set pulse generator for a 10 ns output at an amplitude of 2.00 V.
(11) Use TI CENTER/MARKER FREQUENCY knob to align 2nd marker from left
to TI second vertical graticule line.
(12) Adjust A1A48A1R1065 (fig. 3) for one marker per division over center eight
divisions of CRT (R).
(13) Reduce pulse generator output to minimum and disconnect from TI RF INPUT
50Ω.
13