TB 9-6625-2345-35
(14) Press TI keys as listed in (a) and (b) below:
(a) BLUE SHIFT.
(b) SPAN/DIV, 1, 0, 0, MHz +dBX.
(15) Use TI ◄►POSITION control to realign dot to center graticule of CRT.
(16) Reduce all outputs to minimum and disconnect equipment setup.
b. Adjustments
(1) Press and release TI spectrum analyzer POWER key to OFF.
(2) Disconnect all connections made to TI spectrum analyzer.
(3) Set TI spectrum analyzer on front panel.
(4) Loosen four screws through back rubber feet.
(5) Pull cover up and off.
(6) Place TI on bottom and reconnect power cord.
(7) Depress TI spectrum analyzer POWER key to ON.
(8) Connect equipment as shown in figure 6 below.
(9) Press TI keys as listed in (a) through (g) below:
(a)
BLUE SHIFT.
(b)
RESET.
(c)
BLUE SHIFT.
(d)
SPAN/DIV, 1, 0, MHz +dBX.
(e)
BLUE SHIFT.
(f)
REF LEVEL, 2, 0, kHz dBX.
(g)
TRIGGERING EXT.
17