TB 9-6625-2345-35
Test instrument
Resolution
Marker level
(dBm)
100 Hz
≤ 85
10 Hz
≤ 95
(60) Rotate TI TIME/DIV knob to AUTO.
(61) Press TI keys as listed in (a) through (j) below:
(a) BLUE SHIFT.
(b) RESET.
(c) BLUE SHIFT.
(d) FREQ, 5, 0, MHz +dBX.
(e) BLUE SHIFT.
(f) REF LEVEL, 5, 0, kHz dBX.
(g) BLUE SHIFT.
(h) SPAN/DIV, 1, 0, MHz +dBX.
(i) VIDEO FILTER WIDE.
(j) AUTO RESOLN (not illuminated).
(62) Rotate TI RESOLUTION BANDWIDTH knob to 10 kHz.
(63) Rotate TI PEAK AVERAGE knob fully counterclockwise.
(64) Press TI SINGLE SWEEP key (twice). Wait for sweep completion.
(65) Press TI TUNE key.
(66) Using TI CENTER/MARKER FREQUENCY knob, place marker on highest
point of trace (ignore zero Hz response).
(67) TI MKR LEVEL indication will be within limits specified in first row of table
(68) Press TI keys as listed in (a) through (c) below:
(a) BLUE SHIFT.
(b) FREQ, 1, 5, 0, MHz + dBX.
(c) SWEEP (wait for sweep completion).
(69) Using TI CENTER/MARKER FREQUENCY knob, place marker on highest
point of trace.
(70) TI MKR LEVEL indication will be within limits specified in 150 MHz row of
(71) Repeat (68) and (69) above for remaining frequencies listed in table 25. TI MKR
LEVEL indication will be within
limits specified in
table 25.
39