TB 9-6625-2345-35
(c) BLUE SHIFT.
(d) FREQ, 1, 1, 0, MHz +dBX.
(e) BLUE SHIFT.
(f) REF LEVEL, 3, 0, kHz dBX.
(g) BLUE SHIFT.
(h) SPAN/DIV, 5, 0, 0, kHz dBX.
(i) VIDEO FILTER WIDE.
(j) MIN DISTORTION (not illuminated).
(3) Rotate TI PEAK AVERAGE knob fully counterclockwise.
of 30 dBm.
(5) Adjust signal generator #1 output amplitude to position displayed trace at TI
reference level (top) graticule line.
(6) Press TI keys as listed in (a) through (g) below:
(a) SINGLE SWEEP (twice).
(b) GREEN SHIFT.
(c) STEP SIZE.
(d) GREEN SHIFT.
(e) PEAK FIND.
(f) +STEP.
(g) SINGLE SWEEP.
(7) TI MKR LEVEL indications will be within limits specified in first row of table 26.
(8) Repeat technique of (6) (f) and (6) (g) above for remaining harmonics in table 26.
TI MKR LEVEL indications will be within limits specified in table 26.
MKR LEVEL indication limit
(dBm)
2
≤ 90
nd
3rd
≤ 90
(9) Reduce all outputs to minimum.
(10) Connect equipment as shown in figure 10 below.