TB 9-6625-2345-35
b. Adjustments. None
27. Tracking
Generator Spurious Signals
a. Performance Checks
(1) Connect equipment as shown in figure 12 below.
(2) Adjust TI MANUAL SCAN knob to midrange.
(3) Press TI keys as listed in (a) through (e) below:
(a) BLUE SHIFT.
(b) RESET.
(c) BLUE SHIFT.
(d) FREQ, 1, 0, 0, MHz +dBX.
(e) ZERO SPAN (illuminated).
(4) Press spectrum analyzer keys as listed in (a) through (n below:
(a)
Preset.
(b)
(c)
SPAN, 2, 0, MHz.
(d)
BW/Avg, Res BW, 3, 0, 0, kHz.
(e)
Video BW, 1, 0, kHz.
(f)
AMPLITUDE, Ref Level, 0, dBm.
Scale/Div], 1, 0, dB.
(g)
(h)
(i)
Center Freq.