TB 9-6625-2347-35
Test instrument
DELAY TIME
SEC/DIV
Nonintensified limit
setting
setting
(divisions)
0.5 s
2 s
<5
10 s
50 s
<5
1 ms
<4
0.2 ms
(33) Position TI controls as listed in (a) through (d) below:
(a) CH1 AC GND DC switch to DC.
(b) CH 1 VOLTS/DIV switch to 20m.
(c) SEC/DIV to X-Y.
(d) HORIZONTAL MODE NO DLY INTENS DLY'D switch to NO DLY.
(34) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to TI CH 1 OR
X input.
(35) Set oscilloscope calibrator for a CHAN 1, VOLT mode output of 100 mV at 1
kHz.
(36) Rotate oscilloscope calibrator knob located below EDIT FIELD pushbutton until
displayed trace is exactly 5 divisions in length. If oscilloscope calibrator err display is not
within 5.0%, perform b below.
(37) Set oscilloscope calibrator output to minimum and insert a 50 Ω feedthrough
termination into equipment connection.
(38) Set oscilloscope calibrator for a CHAN 1, LEVEL SINE mode output of 50 kHz
and adjust amplitude for exactly 6 divisions of horizontal display on TI crt.
(39) Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to increase
(40) Oscilloscope calibrator frequency indication will be ≥ 2 MHz.
(41) Set oscilloscope calibrator output to minimum and disconnect equipment setup.
b. Adjustments
(1) Disconnect equipment setup.
(2) Position TI controls as listed in (a) through (r) below:
CH 1 ⇕ POSITION to midrange.
(a)
(b)
VERTICAL MODE CH 1 BOTH CH 2 to CH 1.
(c)
CH 1 and CH 2 VOLTS/DIV switches to .5.
(d)
CH 1 and CH 2 VOLTS/DIV CAL fully cw to detent.
(e)
CH 2 INVERT pushbutton to out position.
(f)
CH 1 AC GND DC switches to DC.
(g)
CH 2 AC GND DC switches to GND.
⇦POSITION⇨ to midrange.
(h)
(i)
HORIZONTAL MODE switch to NO DLY.
(j)
SEC/DIV switch to .1 ms.
(k)
SEC/DIV variable fully cw to detent and pushed in.
(l)
MULTIPLIER knob fully ccw to <X1.
17