TB 9-6625-2349-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument
Identification.
This bulletin provides instructions for the
calibration of Multimeter Mercer, Model 83K. The manufacturer's manual was used as the
prime data source in compiling these instructions. The equipment being calibrated will be
referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms,
Records, and
Reports. Forms, records, and
reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
Dc voltage
Range: 250 mV to 1000 V in 7 ranges
Accuracy: 3% FS
Dc current
Range: 50 A to 500 mA in 5 ranges
Accuracy: 3% FS
Ac voltage
Range: 2.5 to 1000 V in 6 ranges at 400 Hz
Accuracy: 4% FS
Range: 0 to 20 MΩ in 4 ranges (center scale at 20 Ω, 200 Ω, 20 kΩ, 200 kΩ)
Accuracy: 3 of arc
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this