TB 9-6625-2349-35
(3) Set calibrator initial output for 250 mV. Adjust calibrator output controls for a
0.25 V dc TI indication. Calibrator control Error display will indicate within +3 percent.
(4) Repeat technique of (2) and (3) above for TI function/range switch and calibrator
output settings listed in table 3. Calibrator control Error display will indicate within
+3 percent for each TI indication.
Test instrument
Calibrator
function/range
initial output
Test instrument
switch settings
settings
indications
(DCV)
(V)
(V dc)
2.5
2.5
2.5
10
10
10
50
50
50
250
250
250
500
500
500
1 k1
1000
1000
1Connect
calibrator output to TI DC 1k V and COM.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator output to TI + and COM.
(2) Set TI function/range switch to 50 A.
(3) Set calibrator initial output for 50 A. Adjust calibrator output controls for a 50 A
indication on TI. Calibrator control Error display will indicate within +3 percent.
(4) Repeat technique of (2) and (3) above for TI function/range switch and calibrator
initial output settings listed in table 4. Calibrator control Error display will indicate
within +3 percent for each TI indication.
Test instrument
Calibrator
function/range
initial output
Test instrument
switch settings
settings
indications
(DC mA)
(DC mA)
(DC mA)
.5
.5
.5
5
5
5
50
50
50
500
500
500
b. Adjustments. No adjustments can be made.