TB 9-6625-2352-35
Table 7a. Simpson, Model 260-4 Specifications
Test instrument parameter
Performance specification
Dc voltage
Range: 0 to 5000 V
Accuracy: 2% of FS
Ac voltage
Range: 0 to 5000 V
Accuracy: 3% of FS
Range: 0 to 20 MΩ
Accuracy: 3 of arc
Dc current
Range: 0 to 10 A
Accuracy: 2% of FS
Table 7b. Simpson, Model 260-4 Dc Voltage
Dc Voltage
Test instrument1
Calibrator output
Test instrument
Dc volts
Initial
Limits
range
Indication
output
Min
Max
Adjustments
250
250
250
mV
245
mV
255
mV
2
2.5
2.5
2.5 V
2.45 V
2.55 V
10
10
10
V
9.8 V
10.2 V
50
50
50
V
49
V
51
V
None
250
250
250
V
245
V
255
V
1000
1000
1000
V
980
V
1020
V
50003
5000 (900)
5000 (900) V
4900 (800) V
5100 (1000) V
TI must be calibrated in horizontal position.
1
Connect TI positive lead to TI +50 A/250mV input. After 250 mV check, move TI positive lead to TI + input.
2
3 Replace calibrator with power supply. Move positive lead to TI D.C. 5000 V input. Calibrate at 900 V if power supply not
available. After 5000 V check is complete, reduce output to minimum and move positive lead to TI + input.
14