TB 9-6625-2352-35
Table 8a. Simpson, Model 260-5 Specifications
Test instrument parameter
Performance specification
Dc voltage
Range: 0 to 5000 V
2% of FS 0 to 1000 V range
Accuracy:
3% of FS 5000 V range
Ac voltage
Range: 0 to 1000 V
3% of FS 0 to 1000 V range
Accuracy:
4% of FS 5000 V range
Range: 0 to 20 MΩ
2.5 of arc for RX1 range, 2 of arc for
Accuracy:
all other ranges
Dc current
Range: 0 to 10A
1.5% of FS for 50 A range,
Accuracy:
2% of FS for all other ranges
Table 8b. Simpson, Model 260-5 Dc Voltage
Dc Voltage
Test instrument
Calibrator output
Test instrument
1
Dc volts
Initial
Limits
range
Indication
output
Min
Max
Adjustments
250
250
250
mV
245
mV
255
mV
2
2.5
2.5
2.5 V
2.45 V
2.55 V
10
10
10 V
9.8 V
10.2 V
50
50
50 V
49
V
51
V
None
250
250
250 V
245
V
255
V
1000
1000
1000 V
980
V
1020
V
5000
5000 (900)
5000 (900) V
4850 (750) V
5150 (1050) V
3
TI must be calibrated in horizontal position.
1
Connect TI positive lead to TI +50 A/250 mV input. After 250 mV check, move TI positive lead to TI + input.
2
3 Replace calibrator with power supply. Move positive lead to TI D.C. 5000 V input. Calibrate at 900 V if power supply not
available. After 5000 V check is complete, reduce output to minimum, replace power supply with calibrator and move positive
lead to TI + input.
16