TB 9-6625-2356-35
(7) Adjust R8 (fig. 2) for in tolerance reading on multimeter no. 1 (R).
(8) Repeat steps (6) and (7) until readings are in tolerance.
(9) Move temperature selector to the lowest output position.
(10) Adjust R18 (fig. 2) for in tolerance reading on multimeter no. 1 (R).
(11) Repeat steps (6) through (10) until readings are in tolerance.
(12) Move temperature selector to the second lowest output position.
(13) Energize TI by moving the slide switch to the left white ON position.
(14) Adjust R6 (fig. 2) for in tolerance reading on multimeter no. 1 (R).
(15) Repeat steps (4) through (14) above until readings are in tolerance.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.