TB 9-6625-2361-35
(8) Set calibrator for a -20 V dc output. Record absolute value of TI indication.
(9) Set calibrator for a +20 V dc output.
(10) Adjust R32 BALANCE GAIN to the absolute value recorded in (8) above (R).
(11) Repeat (8) through (10) above until the difference between the absolute values is
less than 10 digits without adjustments.
(12) Adjust R45 AC ZERO for a TI indication of 0.00250 V (R).
(13) Set calibrator for 500 Hz and 2.5 V output. Adjust R14 for a TI indication of 2.50000 V (R).
(14) Set calibrator for 50 kHz and 2.5 V output. Adjust C9 for a TI indication of 2.50000 V (R).
(15) Set calibrator for 1 MHz and 1.0 V output. Adjust R75 for a TI indication of 1.00000 V (R).
(16) Set calibrator for 1 MHz and 10.0 V output. Adjust R75 for a TI indication of 10.0000 V (R).
NOTE
Adjust R75 for best in-tolerance condition for (15) and (16)
above.
(17) Set calibrator for 50 kHz and 20.0 V output. Adjust C11 for a TI indication of 20.0000 V (R).
(18) Set calibrator for 500 Hz and 20.0 V output. Adjust R16 for a TI indication of 20.0000 V (R).
(19) Set calibrator for 500 Hz and 160.0 V output. Adjust R18 for a TI indication of 160.000 V (R).
(20) Set calibrator for 50 kHz and 160.0 V output. Adjust C14 for a TI indication of 160.000 V (R).
(21) Set calibrator for 500 Hz and 500.0 V output. Adjust R64 for a TI indication of 500.00 V (R).
(22) Set calibrator for 50 kHz and 500.0 V output. Adjust C4 for a TI indication of 500.00 V (R).
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI INPUT VOLTS/Ω SENSE HI
and LO.
(2) Connect calibrator SENSE HI and LO to TI INPUT AMPS/Ω SOURCE HI
and LO.
(3) Press TI FUNCTION OHMS pushbutton.
(4) Press RANGE UP or DOWN pushbutton to select 10 Ω range.
(5) Set calibrator for a 0 Ω output, EX SNS on.
(6) Press TI ZERO V DC/Ω pushbutton (ZERO annunciator off).
(7) Allow TI to take at least one measurement then press ZERO V DC/Ω
pushbutton (ZERO annunciator on).
(8) Set calibrator for a 10 Ω output then, using output adjustment controls, set
calibrator control display Reading equal to TI indication. If calibrator control display
Error does not indicate within 0.03%, perform b (1) below.