TB 9-6625-2364-24
(c) [Length] use ⇐,⇒,⇑ and ⇓ keys set filter length to 512 in pop up window
then press [Enter].
(d) Press TI Zero Cal, [Zero A] keys and wait for Please Wait symbol to disappear.
(8) TI reading will be 0 0.05 W.
(9) Set meter calibrator output to 144.93 mV and press OPR/STDBY to OPERATE
(10)Press TI [Cal►], [Cal A] keys.
(11)Set TI filter settings to values listed in table 3 using technique of (7) (a) through
(c) above.
(12) Set meter calibrator levels as listed in table 3. TI display will indicate within
Test instrument
Meter calibrator
Test
display indications
output (+DCV)
instrument
filter
settings
Min
Max
256
458.31 V
3.0977 W
W
3.2269
1.4493 mV
256
W
W
9.904
10.096
4.5831 mV
256
W
W
31.428
31.819
14.493 mV
64
W
W
99.49
100.51
45.831 mV
64
W
W
314.72
317.74
144.93 mV
16
0.9954 mW
1.0047
mW
459.12 mV
16
3.1477 mW
3.1769
mW
1.4588 V
16
9.954
mW
10.046
mW
4.6872 V
16
31.477
mW
31.769
mW
15.588 V
16
99.54
mW
100.46
mW
(13) Set meter calibrator output to STANDBY.
(14) Press TI POWER switch to off.
(15) Move cable connection from channel A power sensor input to channel B power
sensor input.
(16) Set TI POWER switch to on position.
(17) Press Preset Local, [Confirm] keys.
(18) Press TI DISPLAY ↕ key to activate B display.
(19) Repeat technique of (4) (c) through (12) above for channel B.
(20) Set meter calibrator output to STANDBY.
(21) Set TI POWER switch to off and disconnect equipment setup.
b. Adjustments. None.
CHANGE 1